KSV Minimicro System 1G
- Miniature Film Balance for Grazing Angle Applications
PC controlled Langmuir Film balance for conventional analysis of monolayers at the air-water interface and Grazing Angle Incident Beam Applications.
Configuration includes
- The KSV MiniMicro Control Module
- Barrier drive
- Film balance
- Special Grazing Angle trough
Unique Features
- Trough ideal for Grazing Angle incident beam applications
Common features
- Symmetric compression of the monolayer by two inter-linked surface barriers
- Leak proof hydrophilic Delrin barriers
- Form-sintered single-piece, non-porous PTFE trough, surface area 170 cm2,(L195 x W85 x D5) and subphase volume 83 ml
- Subphase
temperature control facilitated by aluminum heating/cooling base-plate
controlled by external circulating water bath (bath sold separately)
- Super-sensitive
film balance with platinum Wilhelmy plate (51066), 100 paper plates
(51066FP) and hook for the paper plates (51066FPH)
- Digital display of surface pressure
- Complete KSV WINLB software for Windows operating systems
- Comprehensive instruction manual and instructional video CD on how to set up the instrument and make basic measurements
- Complete starter kit to make initial measurements (see accessory KSV-LBKIT)
| Specifications |
| Film balance measuring range |
|
0 to 250 mN/m |
| Film balance resolution |
|
0.004 mN/m |
| Barrier compression speed |
|
0.02 to 200 mm/min (symmetric compression) |
| Effective trough dimensions |
|
195 x 85 mm |
| Subphase volume |
|
0.083 l |
|